CGI Brain Hemorrhage AI Receives EU Class IIb Certification

Key PointsCertification lets CGI deploy the hemorrhage software across EU hospitals at scale.
View more
- Certification lets CGI deploy the hemorrhage software across EU hospitals at scale.
- External testing showed acquisition settings can sharply alter false-positive rates.
- Neither retrospective study measured the software's effect on a live clinical workflow.
CGI announced on November 24, 2026 that CGI AI Head Analysis, developed with Helsinki University Hospital and Planmeca, had received official certification as a class IIb medical device under the European Union Medical Device Regulation. The software supports physicians in detecting brain hemorrhages from computed tomography (CT) scans, and CGI said the milestone enables hospitals to deploy it at scale.
The software examines head CT images for non-traumatic, spontaneous intracranial hemorrhages and flags suspected acute blood for rapid physician review. Its intended role includes helping doctors rule out overlooked, potentially life-threatening hemorrhages in demanding settings such as busy emergency rooms. CGI described the certification as confirmation that the product meets strict national and EU standards for safety, quality and performance.
Miikka Korja, the HUS chief physician and associate professor of neurosurgery who led the project, characterized the certification as a historic achievement following years of collaboration. Sonja Pietilä, who leads the solution’s development at CGI, said the certified product can be integrated directly into hospital imaging workflows and clinical decision-making.
Clinicians Remain in the Review Process
CGI’s workflow description has the radiologist assess the images independently before using the AI output as diagnostic support and then comparing the assessment with the AI result. The project applies a humans-in-the-loop approach, with experts engaged from design through output interpretation. It also addresses privacy and security risks associated with diagnostic imaging, analysis and data movement. The result is an additional automated check within clinician-led interpretation rather than a stand-alone diagnostic process.
Validation Defines Performance and Deployment Limits
A Scientific Reports study published on August 15, 2025 describes four base convolutional neural networks and a metamodel developed using 300 head CT scans. Researchers retrospectively evaluated the system on 7,797 consecutive emergency head CT scans from ten Finnish hospitals, including 118 scans showing spontaneous intracranial hemorrhage.
With full post-processing, the researchers reported case-level sensitivity of 89.8% and specificity of 89.5%. The system detected all 78 hemorrhage cases imaged confirmedly or presumably within 12 hours of symptom onset and identified five hemorrhages absent from initial on-call reports. Median analysis time in the study was 6.7 seconds per scan.
Those results carried important qualifications. The validation was retrospective and internal, did not assess usefulness in a live clinical workflow, and found that sensitivity declined as the delay between symptom onset and imaging increased.
A separate external validation published on May 25, 2026 tested 901 retrospective head CT scans from University Hospital Zürich and Kantonsspital St. Gallen, of which 81 contained spontaneous hemorrhages. Using the original imaging data, the solution identified 74 of those 81 hemorrhages, producing overall sensitivity of 91.4% and specificity of 78.4%.
The authors found that CT acquisition parameters materially affected false-positive rates. Specificity in the St. Gallen cohort rose from 47.5% to 74.0% after scans were standardized to a 512-by-512 pixel matrix, lifting overall specificity to 85.5% when combined with the original Zürich data. Two acute subarachnoid hemorrhages were missed in Zürich, meaning the complete early-case detection seen in the Finnish validation did not carry over to that cohort. The external study was also retrospective and could not measure effects on clinical workflow or end-user satisfaction.
Certification gives CGI and its partners a regulatory basis for wider hospital deployment. The remaining implementation question is whether hospitals can reproduce the reported detection performance in routine care: external testing showed that acquisition settings can sharply change false-positive rates, while neither retrospective study measured the software’s effect on a live clinical workflow. Imaging-protocol control, local performance evaluation and clinician oversight therefore remain central to deployment.